Sets the number of events contained in one buffer. It can be any number between 1 and 1023. This parameter should be set via the experiment option in order that the acquisition control software can adjust various interrelated buffer options accordingly. Number of Events per Aggregate (0x1n34, 0x8034, R/W, I) The board has a fixed amount of RAM memory to save the events. The memory is divided into a programmable number of buffer, called "aggregates", whose number of events can be programmed by this register. Bit Description [9:0] Number of events per aggregate. Maximum value is 1023. [31:10] Reserved