Decay Time: time constant of the exponential decay at the output of the charge sensitive preamplifier. This parameter is used for the pole-zero cancellation of the trapezoid (expressed in sample periods) This is the register DPP_DECAY_TIME (M) Decay Time (0x1n68 (channel n), 0x8068 (all channels, write only), R/W, ) ‐ Bit Description [15:0] Exponential Decay Time, corresponds to the time constant of the input tail. The correct setting of this parameter is useful for the pole-zero cancellation of the trapezoid. The decay time is expressed in sampling clock unit (10 ns). [31:16] Reserved